Using the Pump
41
Delivering a Demand Dose
In order to deliver a Demand Dose, the pump must be
programmed to allow them in the Setup menu (where
you will need to program a Lockout time limit and dose
duration) as well as in the Delivery Program menu
(where you will program the amount).
A Demand Dose is initiated by pressing the Demand
Dose button on the side of the pump. A Demand Dose
can be given only when the pump is running.
If the patient tries to deliver a Demand Dose during the
lockout period, the display will show, “No dose allowed.
If the pump is not programmed to allow Demand Doses,
pressing the button will have no eect.
Unlike Automatic Doses, the patient chooses whether
or not to deliver a Demand Dose. If the pump is pro-
grammed to allow demand doses, the patient can choose
to deliver a dose or not, as long as it is not prevented by
any Lockout time.
Press Demand Dose
button; Demand Dose
screen appears
To give the dose, press
Deliver. If you do not
want to give the dose,
press Cancel.
Administer demand
dose of 0.66 ml?
Q
Demand
Dose
<Cancel Deliver>

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-MS 3 and is the answer not in the manual?

Smiths Medical CADD-MS 3 Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-MS 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals