Programming the Pump
26
Setup Menu - Delivery
Before programming Delivery for a new patient it is
recommended that you rst use the New Patient feature
(as shown on page 25).
In Delivery you will decide which of the three types of
delivery methods (Continuous Rate, Automatic Dose
and Demand Dose) will be used for the patient, and
program delivery Maximums, Dose Duration and Dose
Lockout Time.
Note: If the pump has already been programmed, you
will not be able to set pump values less than what is
already programmed.
Dose Duration is the time over which a dose (Automatic
and/or Demand) is delivered. You can choose to have
the dose delivered over 1 to 15 minutes (programmed in
1 minute increments). e default value is 12 minutes.
e Dose Lockout time is the minimum amount of
time that can pass between the start of one dose and the
start of the next dose. You can choose a lockout time
of 15 minutes to 24 hours (programmed in 15 minute
increments). e default value is 1 hour. Dose Lockout
time cannot be 0.
Delivery is only available when the pump is stopped.
Note: Dose Duration and Dose Lockout time cannot be
equal.