Programming the Pump
16
Before using the pump for the
rst time
Inserting a battery, programming the regional settings (if
required), and setting the time and date are the rst things you
need to do when you get the pump. e expected battery life is
approximately 2 weeks (battery low alert) at 0.124 ml/hr (the
Vibrate alert uses up battery power faster).
Insert a battery
1. Put the side of a smooth-edged coin into
the slot on the battery cap and turn it
counterclockwise (le) to open it (see
Figure 3). Remove the old battery (if
present). Discard used batteries according
to local laws and requirements.
Warning: ere are potential health
hazards associated with improper disposal of batteries,
electronics, and contaminated (used) infusion sets and
cartridges. Dispose of used batteries, infusion sets, car-
tridges, and other used accessories, or a pump that has
reached the end of its useful life, in an environmentally
safe manner, and according to any regulations that may
apply.
2. Insert one AAA (IEC LR03) alkaline
battery into the compartment, making
sure the + side goes in rst (see
Figure 4); if you insert the battery
wrong, the pump will not start.
Caution: Do not use NiCd, nickel
metal hydride, carbon zinc (heavy
duty), lithium or any rechargeable
batteries. ey will not power the
pump properly, and the battery life
indicator on the home screen may not
show the correct amount.
3. Place the battery cap back over the compartment. Push
down and turn the cap clockwise (right). Again use a
smooth-edged coin to tighten the cap.
Warning: Make sure that the battery cap is fully tight-
ened to avoid an interruption in battery power which
can cause the pump to power down and stop the deliv-
ery of drug therapy. A prolonged interruption in the
delivery of drug therapy can result in serious patient
injury or death.
2002-09-16 D. Zurn
«cozmo-turn-batt-cover.eps»
Figure 3
Figure 4
D. Zurn, 2001-08-29
needle-attach.ai
2002-09-16 D. Zurn
«cozmo-insert-AAA.eps»

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-MS 3 and is the answer not in the manual?

Smiths Medical CADD-MS 3 Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-MS 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals