Programming the Pump
18
Self Tests
During the self tests, the pumps internal computer
performs tests on all the major hardware, computer, and
electronic systems.
Warning: If there are any system problems dur-
ing the self tests, the pump will stop the tests and
display an alarm screen letting you know there is a
problem. If this happens, do not use the pump.
Aer installing a new battery, watch the pumps display
and verify the following:
e internal computers soware version appears, as well
as the pump serial number and last error code (if any).
e entire display becomes a darker gray. Look for any
blank or incomplete areas, which indicates a broken
display. e display then briey goes blank; if the entire
display is not blank, it indicates a broken display.
e pumps internal computer tests the major hard-
ware, computer and electronic systems. If there is a
problem with any system, an alarm occurs and you
will not be able to start delivery.
When the self tests are complete, power up ends and the
pump beeps six times.
e rst time you insert a battery and a new pump per-
forms the self tests, you will see the following alert screens:
A screen will appear that reads, Program Defaulted Set
time and date. Program the current time and date by press-
ing
and
to set the highlighted value, then pressing
Next to move through the sequence. When the time and
date are correct, press Done.
Next you will see a screen that reads, Program Defaulted
You must program the pump. Press OK.
After the pump beeps 6 times, a screen that reads,
Cartridge removed Press OK to begin load process”
is displayed. Press OK; the pump goes to the Cartridge
menu. If you are ready to load a cartridge, follow the steps
indicated starting on page 31. If you arent ready to load a
cartridge yet, press Done.
NOTE: e pump has a built-in internal battery that pow-
ers the clock and allows the pump to store the programs
and history (see page 69) which is charged by the AAA
alkaline battery. If your internal battery has not been
charged for awhile, you may see a screen that reads, Pro-
gram Defaulted Setup of cartridge volume required, with
no cartridge present, press Next”. Make sure no cartridge
is loaded into the pump and press Next.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-MS 3 and is the answer not in the manual?

Smiths Medical CADD-MS 3 Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-MS 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals