Technical Information
71
Delivery specications
Main Menu
Setup
Delivery Methods:
Choose to have the following delivery methods available:
Continuous Rate: Yes or No
Default: No
Automatic Dose: Yes or No
Default: No
Demand Dose: Yes or No
Default: No
Maximum Continuous Rate: 0.000 ml/hr to
1.000 ml/hr in 0.002 ml increments
Default: 0.800 ml/hr
Maximum Auto or Demand Dose: 0.000 ml to
1.000 ml in 0.002 ml increments
Default: 1.000 ml
Dose Duration: 1 to 15 minutes in 1 minute
increments
Default: 12 minutes
Dose Lockout Time: 00:15 minutes to 24:00 hours
in 00:15 (15 minute) increments
Default: 01:00 (1 hour)
Time and Date:
Allows user to set time and date
New Patient:
Mark for new patient: Yes or No
Default: No
Clear Delivery: Yes or No
Default: No
Security:
Delivery Program Menu: Yes or No
Default: Yes
Setup Menu: Yes or No
Default: Yes
Alerts:
Low cartridge volume: Alert at 0.050 ml to
0.500 ml in 0.01 ml increments
Default: 0.2 ml
Display site reminder: Yes or No
Default: No

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-MS 3 and is the answer not in the manual?

Smiths Medical CADD-MS 3 Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-MS 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals