Technical Information
70
internal battery maintains pump programming and his-
tory for approximately 30 days. e life of the internal bat-
tery is 10 years minimum under normal use conditions.
Operating conditions:
Temperature: 2°C to 40°C (35.6°F to 104°F)
Humidity: 90% relative humidity (non-condensing)
maximum
Atmospheric pressure: 70 kPa (or 10,000 feet above sea
level) to 106 kPa (10.2 psi to 15.4 psi)
Caution: Do not use the pump in hyperbaric chambers
as they will aect how the pump works and may also cause
damage to the pump.
Transport and storage conditions:
Temperature: –20°C to 60°C (–4°F to 140°F)
Humidity: 90% relative humidity (non-condensing)
maximum
Atmospheric pressure: 70 kPa (or 10,000 feet above sea
level) to 106 kPa (10.2 psi to 15.4 psi)
System delivery accuracy (ml/hr):
± 3% (nominal). At low delivery rates, this accuracy
may not be achieved for short periods. During the total
delivery time, the accuracy averages out.
System denition:
CADD-MS
™
3 Ambulatory Infusion Pump with Smiths
Medical 3 ml Medication Cartridge and a Unomedical
Comfort
™
Infusion Set
Blockage (occlusion) alert:
18 psi ± 5 psi
Dose accuracy at set value of 0.002 ml:
± 15% (measurement error ± 10%)
Dose accuracy at set value of 1.0 ml:
-0.12, -0.88%; Average - 0.53% (measurement error
± 0.0005%)
Maximum volume infused under single fault condition:
Less than 0.02 ml
Delivery rate during dose:
Approximately 0.0001 ml per second minimum,
0.016667 ml per second maximum (based on dose
amount and dose duration - dose duration congurable
from 1 to 15 minutes)
Delivery rate during Fill tubing/Fill cannula:
Approximately 0.01 ml per second
Continuous Rate Delivery Interval:
Every 3 minutes (approximately
1
/
20
continuous rate)