Page #53 background imageLoading...
Page #53 background image
Using the Pump
47
Continuous Rate Menu
If the pump is programmed with a continuous rate, a
Continuous Rate Home screen is added. A Continuous
Rate Menu is also added, where you can view (but not
change) the Continuous Rate schedule.
Start at
Continuous Rate
(CR)
Home screen
Press
CR Menu
1
6
CR Menu
Menu>
Continuous Rate
0.3 ml/hr
2.988ml
##
CR
o
<Home Select>
Stop Delivery
Review Rates
Continuous
Rate Menu
Choose Review
Rates, then press
Select
p
Home>
Time ml/hr
12:00AM 0.5
06:00AM 0.3
Cont. Rate
Total 8.4 ml
OK
Press
Home
(Press
Home)
(Press
OK)
Press the
or
keys to view the entire
Continuous Rate schedule (if needed).
Press Home to return directly to the
home screen or press OK to return to the
Continuous Rate Menu.