Help
54
Alerts
Message in display What it means / Alarm How to make it stop
e amount of medication le in
the cartridge is low. Alert repeats at
intervals until Cartridge Very Low
alarm occurs. Attention alarm.
Press OK. Replace the cartridge as soon as practical.
An alarm or alert occurred while
you were programming/editing a
value. Single alarm.
Press OK. Reprogram the value.
e cartridge volume needs to be
rechecked to make sure it will be
displayed correctly on the home
screen. Attention alarm.
Remove the cartridge (if present), then press Next.
e time and date settings in the
program have been reset and must
be reprogrammed. Attention alarm.
Press the
or
key to choose the currently high-
lighted setting, then press Next to move highlight to
next setting. When time and date are correct, press
Done.
Cartridge
Volume Low
Change cartridge
soon.
Edit not
saved
Editing was not saved
because an alarm
occurred.
Program
Defaulted
Setup of cartridge
volume required.
With no cartridge
present, press Next.
Program
Defaulted
Set time and date:
XX:XX XX
XX/XX/XX