Using the Pump
36
3. Fill the tubing
Filling forces medication from the cartridge and pushes
air out of the tubing. Filling is complete when you see
medication come out of the end of the tubing and all
air is removed. e amount of medication used to ll
the tubing is not counted as medication delivered to the
patient.
Warning: Never use Fill Tubing when the infusion
set is connected to the body, or an unintended dose
could be delivered.
Press Select
<Back Start
0
Fill
Tubing
ml
<Done Stop
0.09
Filling
Tubing
ml
Press Start
<Back Start
0.25
Fill
Tubing
ml
Tubing begins to ll. When tubing is lled,
press Stop or Done (lling automatically
stops at 0.3 ml)
Press Back
<Back Select>
Continue to ll
tubing?
No
Choose Yes to continue
lling then press Select
NOTE: once the () symbol
appears next to Fill tubing,
you will not be able to re-enter
that step unless you rst press
Done to return to the Main
Menu, then re-enter the Load
menu
- OR - Press Start
again to continue
lling
- OR - With No displayed press
Select if you do not want to
continue lling
<Done Select>
Load
Fill Tubing
Fill Cannula
Reminder
Cartridge
Menu
<Done Select>
Load
Fill Tubing
Fill Cannula
Reminder
Cartridge
Menu

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-MS 3 and is the answer not in the manual?

Smiths Medical CADD-MS 3 Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-MS 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals