Technical Information
72
Regional Settings:
Time Format: 12-hour or 24-hour
Default: 12-hour
Date Format: mm/dd/yy or dd/mm/yy
Default: mm/dd/yy
Decimal Symbol: Decimal shown as Period or
Comma
Default: Period
Delivery Program
Delivery Methods: Choose to use the following delivery
methods:
Continuous Rate: Yes or No
Default: No
Automatic Dose: Yes or No
Default: No
Demand Dose: Yes or No
Default: No
Continuous Rate Schedule:
Programmable (with a possible 48 time/rate segments)
Time: 12:00 to 11:30 in 00:30 increments or
00:00 to 23:30 in 00:30 increments
Default: --:--
Rate: 0.000 ml/hr to Continuous Rate Maximum
(up to 1.000 ml/hr) in 0.002 ml increments
Default: 0.000 ml/hr
Automatic Dose Schedule:
Programmable (with a possible 24 time/dose segments)
Time: 12:00 to 11:30 in 00:30 increments or
00:00 to 23:30 in 00:30 increments
Default: --:--
Dose: 0.000 ml to Maximum Dose (up to 1.000 ml)
in 0.002 ml increments
Default: 0.000 ml
Demand Dose: 0.000 ml to Maximum Dose (up to
1.000 ml) in 0.002 ml increments
Default: 0.000 ml
Load
User loads cartridge, lls tubing, lls cannula, and selects
site change alarm (if displayed)
Fill tubing: Fills in 0.01 ml increments until Stop or
Done is pressed (automatically stops at 0.3 ml).
Fill cannula: 0.000 ml to 0.012 ml in 0.001 ml
increments
Default: 0.000 ml initially, then last programmed
amount