Technical Information
69
Specications
General specications (nominal)
Minimum increment of resolution:
0.005 ml per activation
Size:
3.2 in × 1.8 in × 0.95 in (80 mm × 47 mm × 24 mm)
Weight:
Approximately 3.2 oz (90 g) including battery and
cartridge
Classication:
Type CF
(protection from electric shock)
Moisture protection:
IPX8 (watertight to a depth of 12 [3.64 m] for 3 min,
or 8 [2.4 m] for 30 min.
Pump alarms/alerts:
Pump stopped, blockage, low battery, dead battery, low
cartridge volume, empty cartridge, system fault
Optional alarms/alerts: site change
Maximum infusion pressure:
23 psi
Maximum time to blockage (occlusion) alarm:
Minimum Maximum
Intermediate rate: 0.124 ml/hr 12 min 15 min
Minimum rate: 0.002 ml/hr 9.5 hr 19 hr
Bolus volume at blockage (occlusion) release:
Minimum rate: approximately 0.04 ml
Intermediate rate: approximately 0.21 ml
Power source:
One AAA (IEC LR03) alkaline battery.
Caution: Do not use NiCd, nickel metal hydride,
carbon zinc (heavy duty), lithium or any rechargeable
batteries. ey will not power the pump properly, and the
battery life indicator on the home screen may not show
the correct amount.
Battery life:
e expected battery life is approximately 2 weeks (bat-
tery low alert) at 0.124 ml/hr
Data storage time:
An internal battery powers the clock, and allows pump to
store program parameters and history. e internal battery
is charged by the AAA alkaline battery. Once fully charged
(approximately 40 hours aer AAA battery installed), the