Programming the Pump
27
Starting at
Setup Menu
screen
Press Back
Press Home
Choose Delivery,
then press Select
1
7
k
CR Menu
Menu>
PUMP STOPPED
##A
2.988 ml
CR
Choose Yes or
No then press
Next
Choose Yes or
No then press
Next
Choose Yes or
No then press
Next
k
<Back Next>
Max Continuous
Rate
0.8
ml/hr
Choose 0 to
1 ml/hr then
press Next
k
<Back Next>
Max Automatic
or Demand Dose
1
ml
Choose 0 to 1 ml
then press Next
k
<Back Next>
Dose Duration
12
minute(s)
Choose 1 to
15 minutes
then press Next
k
<Back Next>
Dose Lockout
Time
01:00
hh:mm
Choose 00:15 minutes
to 24:00 hours then
press Next *
o
<Back Select>
Delivery
Time and Date
New Patient
Security
Alerts
Setup Menu
p
<Back Select>
Delivery
Time and Date
New Patient
Security
Alerts
Setup Menu
m
<Home Select>
Start Delivery
Setup
Delivery Program
Load
History
Main Menu
<Back Next>
Enable Delivery
Methods
k
Continuous Rate: No
Automatic Dose: No
Demand Dose: No
<Back Next>
Enable Delivery
Methods
k
Continuous Rate:: Yes
Automatic Dose: No
Demand Dose: No
<Back Next>
Enable Delivery
Methods
k
Continuous Rate: Yes
Automatic Dose: Yes
Demand Dose: No
* You cannot program a Dose Lockout that will cause
one programmed Automatic Dose to lockout another.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-MS 3 and is the answer not in the manual?

Smiths Medical CADD-MS 3 Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-MS 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals