Help
62
Warning: If the pump is dropped or hit against
something hard, it must always be inspected care-
fully to make sure it is still working properly. Make
sure the display is working correctly, and the car-
tridge, cartridge cap, battery cap and infusion set are
connected correctly. If there is damage to the pump
outer shell (cracks, chips), the pump may no longer
be watertight.
Warning: If the display has missing or incomplete
characters, or if the pump does not seem to be
working correctly, stop using the pump immediate-
ly. Contact Smiths Medical MD, Inc. for information
on servicing the pump.
Warning: Make sure that the battery cap is fully
tightened to avoid an interruption in battery power
which can cause the pump to power down and stop
the delivery of drug therapy. A prolonged interrup-
tion in the delivery of drug therapy can result in
serious patient injury or death.