75
Flow rate from startup
Time interval: 15 minutes
Total time: 1530 minutes
Programmed rate: 0.002 ml/hr
Flow rate error
Programmed rate: 0.002 ml/hr
Average ow rate: 0.0021 ml/hr
Mean ow error: 2.59%
-40
-30
-20
-10
0
10
20
30
40
T (min)
60 150
330 570
930
Trumpet Curve
Epmax
Epmin
% Error
of Flow
2005-Nov-28 D. Zurn
«Graph Trumpet MS3 2µl.eps»