redPOWER QUBE Cased Fiber Laser User Manual
104
o Triggering the Alarm Reset function while the state is in the STANDBY or ON
state will switch the state to the OFF state.
o An output function can be configured on the Machine Interface to indicate when
there is an alarm condition.
o This function is only operational in Remote Control.
o This function can be assigned to any of the seven inputs.
o Associated functions
Alarm - Output Function
Laser STANDBY - Output Function
Sequence diagrams
Figure 62 Alarm Reset
Notes:
Alarm condition causes the state to move to the OFF state.
The alarm condition must be cleared before resetting it otherwise the alarm may reoccur
following the reset.