redPOWER QUBE Cased Fiber Laser User Manual
81
9.5.2 Laser Temperature Set Point for Cooling water Control Valve
The operating set point for the solenoid valve can be adjusted using the Solenoid Settings
screen that is accessed from the main Menu bar by selecting Configuration - Laser - Solenoid
Settings. The dialog box, Figure 35, shows the current Set-point value, and allows the
temperature set-point to be updated. The current real time status of the solenoid valve is
reported.
When the laser is not in operation, the Solenoid Open and Close buttons on the dialog can be
used to set the solenoid status manually. During operation the Solenoid Open button will over-
ride the automatic control of the valve and open it for 5 seconds.
Figure 35 Solenoid Set Point Update Screen
9.5.3 Machine Interface
The Machine Interface functions are described in Section 7.6.5. The configuration of the seven
outputs and seven inputs can be designated and confirmed by selecting either the
Maintenance Tab and then MCIF from its toolbar, or from the Menu bar: View Windows
Maintenance Machine Interface.
To alter the configuration use the Menu bar in the Maintenance - Machine Interface window
and select Action Change Configuration. To write the new configuration to the QUBE Fiber
Laser, select Action Save Configuration.
The Function for each connection can be selected from the drop down boxes associated with
each one.