5-10
10083002 Operator Manual Operating Instructions
Sterilization Recommendation
1. Metal handle and silicone ring must be sterilized before use in
any surgical procedure.
2. When placing handle and ring in the sterilizer for processing, the
best practice is to orient all objects with openings downward, as
if to allow water (or air) to flow out.
3. Sterilize handle and silicone ring separately (disassembled, but
in the same chamber) using a cycle with either of the following
parameters:
Prevacuum
Temperature: 270°F (132°C)
Exposure Time: 4 minutes
Gravity
Temperature: 270°F (132°C)
Exposure Time: 10 minutes