11-1
Appendix - EMC Compliance Technical Data Operator Manual 10083002
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions for all ME Equipment and
ME Systems
Table 11-1. Guidance and Manufacturer’s Declaration – Electromagnetic Emissions – for all ME Equipment
and ME Systems (per IEC 60601-1-2 clause 5.2.2.1 c) Table 1)
The HarmonyAIR Surgical Lighting System M Series with Harmony LED Wall Control Unit with or
without video monitor and video-monitor related components is intended for use in the
electromagnetic environment specified below. The Customer or the user of the HarmonyAIR
Surgical Lighting System M Series with Harmony LED Wall Control Unit should assure that it is
used in such an environment.
Emissions test Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The HarmonyAIR Surgical Lighting System M Series with
Harmony LED Wall Control Unit uses RF energy only for
its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A
The HarmonyAIR Surgical Lighting System M Series with
Harmony LED Wall Control Unit is suitable for use in all
establishments, other than domestic and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions IEC
61000-3-3
Complies
APPENDIX – EMC COMPLIANCE
TECHNICAL DATA
11