6-2
10083002 Operator Manual Cleaning the Equipment
6.2 General Cleaning/
Disinfecting Procedure
NOTE: The user must follow the requirements of the national
committee responsible for hygiene and disinfection.
1. Ensure system is turned off and lightheads are cooled
sufficiently.
2. Wear rubber gloves.
3. Prepare an approved cleaning or disinfecting solution in
accordance with the directions on the container’s label.
4. DO NOT SPRAY anything directly onto any system components.
Dampen a soft cloth with the cleaning solution and wring out the
excess moisture.
5. Thoroughly wipe the areas to be cleaned (See Figure 6-1).
6. Rinse all surfaces with a clean soft cloth wipe and clear water.
7. Wipe all surfaces dry with a clean dry soft cloth.
8. Verify metal sterilizable handle (when used) is sterilized using a
standard hospital cycle.
NOTE: Always sterilize metal sterilizable handle between
procedures. Sterilize lighthandle using conventional hospital
sterilization procedures and a standard sterilization prevacuum
or gravity cycle.
Important: If disposable lighthandle covers are used, verify a new
one is used for each procedure.
WARNING – PERSONAL
INJURY HAZARD: Do not
attempt to clean lighthead
unless power is turned off and
the lighthead has cooled
sufficiently.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE
HAZARD:
Use of any disinfectant
solution OTHER than those
listed here may cause
discoloration or deformation
on the lens surface:
Coverage
®
Plus Germicidal
Surface Wipes Disinfecting/
Deodorizing/Cleaning Wipes.
Cleaning solutions other than
those listed have NOT been
tested for compatibility or
effectiveness. Always follow
manufacturer instructions for
concentrations and use of
cleaning products.
Avoid discoloration of control
center keypad and display. Do
not clean control center with
povidone-iodine solutions or
allow such solutions to
contact keypad and display
surfaces.
Prevent leakage of fluids into
interior of lighthead or control
center.
Cleaning and disinfecting
agents used on this lighting
system must be certified by
their manufacturer to be
compatible with the following
materials: polycarbonate,
polyetherimide, santoprene.
DO NOT SPRAY any cleaning
product directly onto the
lighthead, modem chassis, or
any system components.
Dampen a soft cloth with the
cleaning solution and wring
out the excess moisture.
!
!