Installation and commissioning
Instructions for use • AIDA C WD310 • AAB272_EN_V2.0_12-2021_IFU_CE-MDR 30
Message Status Description Action
Contact the
DICOM
administrator.
CommitTime-
out
Warning Commit requests were sent, but no
answer has yet been received from
the Commitment server for at least
one object. The first request was
made at least as long ago as the
configured timeout.
If failed and timed-out objects are
present, "Timeout" is set.
Re-send the
Commit request.
CommitPend-
ing
Normal Commit requests were sent, but no
answer has yet been received from
the Commitment server. The first
request was made within the con-
figured timeout period.
Wait for
transmission to
take place.
CommitIn-
complete
Normal No Commit request was sent for at
least one object.
Observe the
timeout period,
then re-send the
Commit request.
StorageIn-
complete
Normal At least one object was not stored.
Observe the
timeout period,
then re-send the
Commit request.
StorageNot-
Done
Error Error when storing the objects on
the store server; no Commit re-
quest was sent.
Re-send the
Commit request.
Contact the
DICOM
administrator.
StorageNot-
Configured
Error No store server configured; Com-
mit request not possible.
Check the DICOM
server settings.
6.1.2.7 System
The
System
tab is used to make various settings at operating system level, such as network or
WLAN settings. Local users can be created and free text can be stored via the Windows
desktop.