Electromagnetic compatibility
Instructions for use • AIDA C WD310 • AAB272_EN_V2.0_12-2021_IFU_CE-MDR 98
Interference im-
munity tests
EN/IEC60601 test
level
Compliance level Electromagnetic envi-
ronment – guidelines
Magnetic field at
the power fre-
quency (50/60Hz)
acc. to IEC
61000-4-8
30A/m at 50Hz/
60Hz
30A/m at 50Hz/
60Hz
If image distortion oc-
curs, it may be neces-
sary to install the prod-
uct further from sources
of electromagnetic
fields or to install mag-
netic shielding. Before
the product is installed,
the electromagnetic
field should be mea-
sured to ensure that it
is sufficiently low.
Immunity test acc.
to IEC61000-4–3
for radiated, ra-
dio-frequency
electromagnetic
fields
3V/m 80MHz to
2.7GHz
* Refer to Table 2 for
wireless proximity RF
field test levels
3V/m 80MHz to
2.7GHz
Immunity to con-
ducted distur-
bances, induced
by radio-fre-
quency fields acc.
to IEC 61000-4-6
3V
rms
on 150kHz to
80MHz
1kHz 80% AM modu-
lation
6V
rms
in ISM band
3V
rms
on 150kHz to
80MHz
1kHz 80% AM modu-
lation
6V
rms
in ISM band
10.4 Table 2 – Test levels for proximity fields from HF
wireless communications equipment
Test fre-
quency
MHz
Frequency
band
MHz
Radio service Modulation Immunity
test level
V/m
Compliance
level
V/m
385 380 – 390 TETRA 400 Pulse modula-
tion 18Hz
27 27
450 430 – 470 GMRS 460,
FRS 460
FM ±5kHz
deviation
1kHz sine
wave
28 28
710 704 – 787 LTE band 13
and 17
Pulse modula-
tion
217Hz
9 9
745
780
810 800 – 960 GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE band 5
Pulse modula-
tion
18Hz
28 28
870
930
1720 1700 – 1990 GSM 1800,
CDMA 1900,
GSM 1900,
DECT,
Pulse modula-
tion
217Hz
28 28
1845

Table of Contents