PARTS REPLACEMENT
4 - 4 STRYKER MEDICAL/BERTEC MÉDICAL
Half-Length Safety Side-Rail
Side-rails are not interchangeable. Make sure you have a spare part on
hand that is identical to the one to be replaced.
If the side-rail to be changed is equipped with a fixed pa-
tient control, you will have to replace the whole mechanism
attaching side-rail to sleeping surface.
REPLACING THE ENTIRE SIDE-RAIL MECHANISM
1. Raise section of sleeping surface high enough to gain access to the
underside of the sleeping surface.
To change a foot end safety side-rail, tilt foot and thigh sections
over, as indicated in steps 4 and 5 in Motor section, page 4-7.
2. If the side-rail is equipped with PATIENT CONTROLS:
Movable control: Unscrew ring and disconnect the patient control
wire connecting side-rail to bed frame.
Fixed control: Find the receptacle located under the sleeping
surface central section and disconnect the electric wire attached to
the fixed patient control. Remove protective cover on headboard and
withdraw wire.
3. Remove the 4 machine screws attaching entire side-rail mechanism
to sleeping surface.
4. Place new side-rail in position and tighten down with 4 machine
screws.
5. Movable control: Reconnect patient control electric wire.
Fixed control: Feed patient control wire under sleeping surface.
Reconnect to outlet. Replace protective cover.
6. Adjust sleeping surface to horizontal position.
REPLACING THE UPPER SECTION OF SIDE-RAIL ONLY
Side-rail without fixed patient control only
1. Use a screwdriver to remove plastic caps sealing the 2 axles, or
pierce in middle, break and remove with a sharp tool.
2. Use a hammer and blunt-ended tool to push protection caps out of
other end of axles.
3. Remove retaining rings using a pair of pliers.
4. Remove the upper section and replace with new part, repeating
above instructions in reverse order.