STRYKER MEDICAL/BERTEC MÉDICAL 1 - 9
PREPARATION PRIOR TO USE
Operation
Hospital beds must be properly installed and their operation checked
before they are put in service.
It is imperative that a visual inspection of material be conducted and bed
operation checked to ensure that no damage occurred during shipping.
SWITCHING ON POWER
1. Plug the bed into a 120V-60 Hz electric receptacle.
2. Place main switch (A) in the ON position.
Always plug bed into a three-prong hospital grade receptacle to
ensure that machine is grounded.
CHECKLIST
BED MOBILITY
Lock casters to make sure that they are working properly.
Refer to Moving the Bed section, page 2-6.
SLEEPING SURFACE POSITIONS
Raise and lower the head and thigh sections (B). Make sure they
operate smoothly and return to a horizontal position.
Refer to Bed Elevation and Elevation of Sections of Sleeping Surface section, page 2-3.
HEIGHT OF SLEEPING SECTION
Raise bed to its highest position and lower it to its intermediate
position. Movement should be smooth.
Refer to Bed Elevation and Elevation of Sections of Sleeping Surface section, page 2-3.
TRENDELENBURG POSITIONS
Made sure that the mechanism setting Trendelenburg positions is
operating correctly.
Refer to Trendelenburg and Reverse Trendelenburg Positions section, page 2-4.