2. Place the trolley in the loading position to expose the head end of the anchor.
3. Locate the window plug (002900010000) and note the status of each battery LED (Figure 8).
FFiigguurree 88 WWiinnddooww pplluugg llooccaattiioonn
4. Use the table below to diagnose inductive primary board status.
GGrreeeenn LLEEDD RReedd LLEEDD SSttaattuuss//aaccttiioonn AAccttiioonn
Solid on Off GGoooodd:
Incoming voltage is
sufficient to charge the cot
and trolley batteries.
No action required.
Flashing in short bursts Off DDiiaaggnnoossee ffuurrtthheerr:
Incoming voltage is too low
or the time limited charging
mode ended.
CChheecckk tthhee vvoollttaaggee
a. Remove the anchor
inductive primary cover
(639001010101).
b. Disconnect the vehicle
to anchor cable and
confirm that the voltage
is greater than or equal
to 12.7V.
c. If the voltage is
sufficient, there may be
an issue with the
inductive primary
assembly.
6390-309-002 Rev AB.0 31 EN

Table of Contents

Other manuals for Stryker Power-LOAD 6390

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Power-LOAD 6390 and is the answer not in the manual?

Stryker Power-LOAD 6390 Specifications

General IconGeneral
BrandStryker
ModelPower-LOAD 6390
CategoryMedical Equipment
LanguageEnglish

Related product manuals