EasyManua.ls Logo

Stryker Power-LOAD 6390 - Page 44

Stryker Power-LOAD 6390
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A
B
C
FFiigguurree 1155 MMaannuuaall rreelleeaassee bbuuttttoonn aasssseemmbbllyy
7. Unplug the cables from the master On/Off switch (D) and remove the manual release button assembly (Figure 16).
NNoottee -- Make note of the cable connection locations, so they do not get mixed up.
D
FFiigguurree 1166 OOnn//OOffff sswwiittcchh ccaabbllee llooccaattiioonnss
8. Reverse steps to reinstall.
9. Verify proper operation before you return the product to service.
EN 38 6390-309-002 Rev AB.0