AR 2000 Operator's Manual
33
Table 4.5
Optional Accessory Specifications for
Peltier Concentric Cylinder System
Temperature range
with tank and pump 0°C to 100°C
with plumbed water supply -10°C to 150°C
with fluid at -20°C -40°C to 100°C
Ramp Rate
Cooling 15°C/min maximum
Heating 13°C/min maximum
Pt100 Internal Resolution 0.01 °C
Table 4.6
Optional Accessory Specifications for Upper Heated Plate
Temperature range
plumbed water supply (11°C) 20°C to 150°
low viscosity silicone circulating fluid at -40°C -30°C to 55°C
vortex air cooler -5°C to 150°C
Ramp rate 15°C maximum
Maximum temperature difference between plates 0.1°C