Performance Tests
CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual
4-67
H Use the keypad to set vertical offset to the setting listed in the table
for the current vertical scale and position settings. The baseline level
may move off screen.
H Set the generator to the level and polarity indicated in the table for
the vertical scale, position, and offset settings you have made. The
DC test level may be off screen. For all steps except the ones on the
1 V/div range, adjust the vertical offset setting as needed to bring the
DC test level to +3.8 divisions (0.2 divisions below the top of the
graticule).
e. Measure the test signal: Press Close.
H Read the measurement results at the measurement statistics
measurement readout. S ee Figure 4--10.
H Record the Mean in the Measurement Mean column of Table 4--4.
Turn on the
measurement called
mean and read the
results here.
Figure 4- 10: Measurement of DC gain accuracy
f. Measure second mean:
H Set the generator to the second level and polarity indicated in the
table for the vertical scale, position, and offset settings you have
made. (For some settings it is possible that the DC test level appears
off screen, while the measured mean value is within tolerance.)
H Repeat substep e using the current vertical scale, position, offset, and
new generator setting for the second mean.
g. Check against limits:
H Subtract the second measurement mean from the first measurement
mean for the current vertical scale, position, and offset.