EasyManua.ls Logo

Tektronix CSA7404B - Figure 4- 23: Final Test Hookup

Tektronix CSA7404B
348 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests
4-94
CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual
H Touch the Pk-Pk button.
H Touch the Close button.
H Set the Vertical SCAL E to 50 mV.
H Set the generator for a 10.0 MHz sine wave.
H Set the generator to output a 4 division signal. Adjust the output
until the Pk-Pk readout displays 200 mV.
c. Set the instrument controls:
H Move the cable from the CH 1 input to the rear-panel Ext Ref input
(see Figure 4--23).
H Touch Menu to select menu mode.
H Touch Utilities and select External Signals.
H Touch the Internal button to select the external reference (the button
name changes to External).
50 Ω Coaxial cable
Output
Sine-wave
generator
Instrument under test
Figure 4- 23: Final t est hookup
6. Confirm external refer ence:
a. Perform a signal path compensation:
H Touch Utilities and select Instrument Calibration.
H Touch Calibrate and wait for the signal path compensation to finish.
b. Check the completion status: If the Status is Fail, refer the instrument to
qualified service personnel.

Table of Contents

Other manuals for Tektronix CSA7404B

Related product manuals