EasyManua.ls Logo

Tektronix CSA7404B - Table 4- 11: Word Recognizer Data

Tektronix CSA7404B
348 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests
CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual
4-117
o. Verify that the instrument triggers at the 0 in the input signal (see
Figure 4--36). Enter pass or fail in the test record.
p. Touch the Clear button.
q. Enter data into Serial Pattern Data field for the next setting in Table 4--10
that is not yet checked.
r. Touch Apply.
s. Verify that the instrument triggers one Unit Interval (UI) after the 0 in
the input signal (see Figure 4--36). Enter pass or fail in the test record.
Table 4- 11: Word recognizer data
Serial pattern data Trigger location
4924 9249 2492 4924
16
One UI before the 1
9249 2492 4924 9249
16
At the 1
2492 4924 9249 2492
16
One UI after the 1
3. Verify that the serial path and pattern matching circuits can do isolated 1s:
a. Adjust the trigger LEVEL to trigger at 75% (+1 division) on the sine
wave.
b. Touch the Clear button.
c. Enter data into the Serial Pattern Data field for one of the settings in
Table 4--11 that is not yet checked. (Start with the first setting listed.)
d. Touch Apply.
e. Verify that the instrument triggers one Unit Interval (UI) before the 1 in
the input signal (see Figure 4--37). Enter pass or fail in the test record.
f. Touch the Clear button.
g. Enter data into the Serial Pattern Data field for the next setting in
Table 4--11 that is not yet checked.
h. Touch Apply.
i. Verify that the instrument triggers at the 1 in the input signal (see
Figure 4--37). Enter pass or fail in the test record.
j. Touch the Clear button.
k. Enter data into the Serial Pattern Data field for the next setting in
Table 4--11 that is not yet checked.

Table of Contents

Other manuals for Tektronix CSA7404B

Related product manuals