EasyManuals Logo
Home>Tektronix>Measuring Instruments>CSA7404B

Tektronix CSA7404B Service Manual

Tektronix CSA7404B
348 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #186 background imageLoading...
Page #186 background image
Performance Tests
4-110
CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual
5. Disconnect the hookup: Disconnect the cables and adapters from the inputs
and outputs.
Equipment
required
Two dual-banana connectors (Item 5)
One BNC T connector (Item 6)
Two precision 50 Ω coaxial cables (I tem 4)
One DC calibrat ion generator (Item 9)
One SMA-to-BNC adapter (Item 19)
Prerequisites See page 4--17. Also, the instrument must have passed Check
Accuracy For Long-Term Sample Rate and Delay Time Accuracy and
Reference on page 4--92.
1. Install the test hookup and preset the instrument controls:
a. Hook up test-signal:
H Connect one of the 50 Ω cables to CH 1 through an adapter. See
Figure 4--31.
H Connect the other end of the cable just installed to the PROBE
COMPENSATION output. See Figure 4--31.
BNC cable from PROBE
COMPENSATION output to
CH 1 input
Instrument under test
Figure 4- 31: Initial test hookup
b. Initialize the instrument: Press the DE FAULT SETUP button.
c. Modify the initialized front-panel control settings:
H Set the Vertical SCALE to 100 mV.
Check Probe
Compensation Output

Table of Contents

Other manuals for Tektronix CSA7404B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix CSA7404B and is the answer not in the manual?

Tektronix CSA7404B Specifications

General IconGeneral
BrandTektronix
ModelCSA7404B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals