EasyManua.ls Logo

Tektronix CSA7404B - Page 187

Tektronix CSA7404B
348 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests
CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual
4- 111
H From the tool bar, touch the Vert button and then touch Offset.
Adjust the Ch1 Offset to --0.25 V using the multipurpose knob.
H Set the Horizontal SCALE to 200 s.
H Press PUSH TO SET 50%.
H Use the Vertical POSITION knob to center the display on screen.
H From the tool bar, touch Horiz and select the Acquisition tab.
H Touch Average and set the number of averages to 128.
2. Confirm that the Probe C ompensator signal is w ithin limits for frequency:
a. Measure the frequency of the probe compensation signal:
H From the button bar, touch MEAS and select the Time tab.
H Touch the Freq button.
b. Check against limits:
H CHECK that the CH 1 Freq readout is within 950 Hz to 1.050 kHz,
inclusive. See Figure 4--32.
H Enter the frequency on the test record.
H Touch Clear to remove the measurement.
Figure 4- 32: Measurement of probe compensator frequency

Table of Contents

Other manuals for Tektronix CSA7404B

Related product manuals