Specifications
1-28
CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual
Certifications and Compliances
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility.
Compliance was demonstrated to the following specifications as listed in the
Official Journal of the European Communities:
EN 61326. EMC requirements for Class A electrical equipment for measurement,
control, and laboratory use. Annex D.
12
H IEC 61000--4--2. Electrostatic discharge immunity
H IEC 61000-- 4--3. RF electromagnetic field immunity
3
H IEC 61000--4--4. Electrical fast transient / burst immunity
H IEC 61000--4--5. Power line surge immunity
H IEC 61000--4--6. Conducted RF Immunity
3
H IEC 61000--4--11. Voltage dips and interruptions immunity
EN 61000- 3- 2. AC power line harmonic emissions
EN 61000- 3- 3. Voltage changes, fluctuations, and flicker
Complies with EMC provision of Radiocommunications Act per these stan-
dard(s):
H AS/NZS 2064.1/2. Industrial, Scientific, and Medical Equipment: 1992
1
Emissions which exceed the levels required by this standard may occur when this
equipment is connected to a test object.
2
If interconnect cables are used, they must be low-EMI shielded cables such as the
following Tektronix part numbers or their equivalents: 012-0991-00, 012-0991-01,
012-0991-02 or 012-0991-03 GPIB Cable; 012-1213-00 (or CA part number 0294-9)
RS-232 Cabl e; 012-1214-00 Centronics Cable; or LCOM part number CTL3VGAMM-5 VGA
Cable.
3
Under t hese conditions, the specificat ions are amended as follows:
10 mV/division to 1 V/division: ≤0.4 division waveform displacement or ≤0.8 division
increase in peak-to-peak noise
5 mV/division and 2 mV/division, typical: ≤8 mV increase in peak-to-peak noise
EC Declaration of
Conformity - EMC
Australia / New Zealand
Declaration of Conformity
-EMC