EasyManua.ls Logo

Tektronix CSA7404B - Verify the File System; Figure 4- 6: Setup for the File System Test

Tektronix CSA7404B
348 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Brief Procedures
4-14
CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual
H As you rotate the lower multipurpose knob, the square-wave
probe-compensation signal can become triggered and untriggered.
(Leave the signal triggered.)
c. Verify the delayed trigger counter:
H Double click the Trig Delay control to pop up a numeric keypad for
that control.
H Click the keypad to enter a trigger delay time of 1 second and then
click Enter.
H Verify that the trigger READY indicator on the front panel flashes
about once every second as the waveform is updated on-screen.
8. Remove the test hookup: Disconnect the BNC cable from the channel input
and the probe compensation output.
Equipment
required
One BNC cable, such as Tektronix part number 012-0076-00
One TCA-BNC adapt er
One 1.44 Mbyte, 3.5 inch DOS-compatible formatted disk.
Prerequisites None
1. Initialize the instrument:Push the front-panel DEFAULT S ETUP button.
2. Hook up the signal source: Connect the BNC cable from the probe com-
pensation output to the CH 1 input through a TCA-BNC adapter as shown in
Figure 4--6.
BNC cable from PROBE
COMPENSATION output to
CH 1 input
Instrument under test
Figure 4- 6: Setup for the file system test
3. Insert the test disk: Insert the floppy disk in the floppy disk drive.
4. Set up the instrument: Push the front panel AUTOSET button.
Verify the File System

Table of Contents

Other manuals for Tektronix CSA7404B

Related product manuals