EasyManua.ls Logo

Tektronix CSA7404B - Figure 4- 14: Check of Maximum Input Voltage

Tektronix CSA7404B
348 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests
4-76
CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual
NOTE. When setting the Fluke 9500B to output more than 5 V, use the following
procedure:
Press the Aux button
Press the fourth soft key down (Selects the pulse with an exclamation point)
Set the amplitude to 6.5 V or 10 V
Press the ->| key to select the pulse ener gy
Set the energy to 50J
Press the Output On key
Press the Trig Pulse soft key to trigger the pulse (this will generate a pulse with
59.172, 30.864, and 25 seconds duration respectively).
Use the normal DC output for the 1 V, 3 V, and 5 V generator settings.
Turn on the
measurement called
high and read the
results here
After removing the
10X attenuator, check
the coupling readout
of the channel you are
testing
Figure 4- 14: Check of maximum input voltage
e. Check an unchecked generator setting against limits:
H Remove the 10X attenuator.
H CHECK that the coupling readout on screen for the selected channel
is as listed for the current vertical scale and position/offset/generator
settings. Enter result on test record.
H Reinstall the 10X attenuator.
f. Check the next generator setting: Repeat substeps d and e, using the new
generator setting as is listed in the table.
g. Check the remaining vertical scale settings: Repeat substeps c through f
until all vertical scale settings, listed in Table 4--6, are checked for the
channel under test.

Table of Contents

Other manuals for Tektronix CSA7404B

Related product manuals