TG8000 remote commands
[:EVENt]?
:CONDition?
:ENABle <numeric_value>
:PTRansitio
n
<numeric_va
lue>
:NTRansition <numeric_value>
:PRESet
:QUEue
[:NEXT]
:ENABle
:QUEStio
nable
[:EVENt]?
:CONDition
:ENABle <numeric_value>
:PTRansition <numeric_value>
:NTRansition <numeric_value>
:MAP
<nume
ric_value>,
<numeric_value>
:INS
Trument
[:EVENt]?
:CONDition?
:ENABle <numeric_value>
:PTRansition <numeric_value>
:NTRansition <numeric_value>
Command description
PRESet. The OPERation node covers areas of instrument operational events.
This node reports such items as signal settling, calibration, or running a sequence
QUEStionable.
QUEStionable. The QUEStionable node covers areas of questionable conditions
that exist within the instrument. Questionable conditions might cause you to be
unsure of the quality of the generated signals.
Subnodes. Each node has a set of subnodes that query the contents, or control
the involveme nt, of each bit:
ENABle selects which bits are active in reporting the status.
EVENt is a destructive query only of the status.
CONDition is a nondesctructive query only of the status.
PTRansition allows a status to become true on a positive transition of the
event.
Draft
TG8000 Multiformat Test Signal Generator Programmer Manual 23