HDVG7 HDTV Digital Video Generator m odule remote c ommands
Sets the type of
the serial digital black signals output from the BLACK
connectors. The options are BLACK (serial digital b lack signal), GRAY
(40% flat field signal), FF50P (50% fl at field signal), or WHITe (100% flat
field signal). This command is only available if Option BK is installed.
:OUTPut2:SIGNal?
Returns the current serial digital black signals output from the BLACK
connectors. This command is only available if Option BK is installed.
:OUTPut4:TIMECode:SOURce DIS able|TOD|COUNter
Sets time code source. Disable shuts off the time code output. TOD sets
the time of day clock as the source. Counter sets the time c ode source to
the p rogram counter.
:OUTPut4:TIMECode:SOURce?
Returns the current time code source setting.
:OUTPut4:TIMECode:OFFSet <NR 1>,<NR1>,<NR1>,<NR1>
Sets the time code offset in hours, minutes, seconds, and frames
(HH:MM:SS:FF).
:OUTPut4:TIMECode:LINe <NR1>
Sets the line number on which the VITC is located. For NTSC output, the line
number range is 10-20. For PA L output, the line number range is 6-22. For
HD output, the line number range is 7 -40.
:OUTPut4:TIMECode:ATC:STA Te 0 | 1 (BOOLEAN)
Enables or disables the use of a ncillary time code.
:OUTPut4:TIMECode:ATC:STA Te?
Returns the ancillary time code state. A setting of 0 means ancillary time code
is not used. A setting of 1 means ancillary time code is used.
:OUTPut:CONFigure?
Returns the availability of serial digital black signal output option (Option
BK). "1" indicates that the option is available and "0" indicates that the option
is not available.
SENSE commands
Use these commands to set the timing offset of the SIGNAL outputs and the
BLACK outputs (Option BK only) relative to the intern
al reference signal (frame
reset signal).
Refer to Operating Basics in the TG8000 User Manual for detailed information
about the setting range.
Draft
144 TG8000 Multiformat Test Signal Generator Programmer Manual