45
Screen saver
OFF: no screen saver.
ON: turns into the screen saver mode (displays TELEDYNE
OLDHAM SIMTRONICS logo) if no key is pressed for a certain
period of time.
Averaged value
OFF: averaged gas measurement value is not displayed.
ON: displays the averaged gas measurement values over the
last fifteen minutes or ei
ght hours depending on the settings
done with COM 32. Typically used when toxic gas detectors.
5. Language
Selection of the display menu language.
7.6 Program
1. Buzzer On/Off
Activates or deactivates the internal buzzer of the
MX32v2
.
2. Tag set Allows for the modification of detector tags previously programmed
via
COM 32
.
3. Alarm settings
Allows for the modification of detector alarms previously
programmed via
COM 32
.
4. Port RS485 Configuration of Port RS485 (speed, parity, stop bits, slave number).
This configuration is only useful if the
MX32v2
is equipped with the
RS485 communication card.
7.7 Calibration
If the measurement cell has changed, it is important to declare this through a menu no.
5 Cell change
.
1. Detector select.
This menu enables the selection of detectors to be calibrated (calibration from
MX32v2
or on
the detector).