Section I β CD AutoPlus Maintenance
CD14 Tester & CD AutoPlus Autosampler User Guideβ75-108-800 Rev. Bβ31 May 2021 104 of 130
EAR99 technology subject to restrictions on cover page.
ξ
7. Using the autosampler controls open the valves for
any lines.
8. Gently pull the syringe plunger down and align it with
the drive bar.
9. Reinstall the shoulder screw.
NOTE: The shoulder screw should be able to be
installed finger tight until it is flush with the syringe
plunger. If the screw cannot be turned with only your
fingers, adjust the plunger and try again. Once the
screw is flush with the syringe plunger, it may be
tightened with a screwdriver.
Volume Calibration
Use DI-grade water at room temperature.
1. Weigh empty sample collection vials or test tubes and
record the results (tare weight). Caps are required in
order to ensure accuracy.
2. Program the autosampler to collect the required
sample volume.
3. Weigh full sample collection vials and subtract the
empty (tare) weight from the full weight. Make any
necessary calculations to account for water density.
ππππ’ππ
σ°
ππΏ
σ°
ο
σ°πΉπππ ππππβπ‘
σ°
π
σ°
ο ππππ ππππβπ‘
σ°
π
σ°
σ°
π·πππ ππ‘π¦ σ°
π
ππΏ
σ°
If the volume does not fall within the specifications in the
specifications section, contact Teledyne Hanson Technical
Support.