Section J – Troubleshooting
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021 116 of 130
EAR99 technology subject to restrictions on cover page.
AutoMag Troubleshooting
Problem
Possible
Causes
Recommended Action
Magazine is not
moving the proper
distance
Magazine may
not be
calibrated.
Perform magazine
calibration.
Method not
set up for
proper sample
distance.
Change magazine travel
in method.
Magazine is not
moving at all.
Magazine not
configured.
Ensure method makes
use of magazine.
Ensure magazine is set up
in Device Setup.
Component
failure.
Contact Teledyne Hanson
Technical Support.
Vessel temperature
probes not working.
Vessel
temperature
probes not
configured.
Ensure probes are
configured in the device
setup.
Vessel
temperature
probes not
connected.
Ensure that probes are
connected to tester main
board.
Component
failure.
Contact Teledyne Hanson
Technical Support.
When measuring
temperature, Probe
Error or --.-- is
displayed
Probe is
disconnected.
Reconnect temperature
probe.
Probe is
damaged.
Contact Teledyne Hanson
Technical Support.