Section D – Tester Installation
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021 21 of 130
EAR99 technology subject to restrictions on cover page.
Space Requirements
See the specifications section to determine space
requirements based on the size of the equipment.
The bench should be level and flat.
The bench should be capable of supporting the full weight
(including water) of each machine as stated in the
specifications section.
The CD14 Tester should be placed in the center of the
autosamplers, if used.
Electrical Requirements
Each CD14 dissolution tester requires two grounded outlets
within 3 meters (9 ft.) of the location of the system; one for the
dissolution unit and one for the heater box.
The CD14 Testers are available in one configuration, 100 V
through 240 V. The voltage is automatically set by the
instrument.
Power cords should meet the following specifications: IEC
320-C13 connector; SJT 18AWG3C, 2 x 0.824 mm2, 60 °C,
300 V.