Section H – CD14 Tester Maintenance
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021 97 of 130
EAR99 technology subject to restrictions on cover page.
Section H – CD14 Tester Maintenance
Scheduled Maintenance Overview
Spills
Immediately clean up any spills.
After Each Test
Wipe down the system with a clean damp cloth when finished.
Pay close attention to the vessel plate and clean and dry any
spills.
Clean baskets, basket shafts, paddles and vessels as needed.
Avoid harsh or abrasive materials which may cause damage to
these parts. It is best to dry these items right away, so they
don’t stain.
Weekly
Check bath water level and fill if needed.
Inspect bath water for algae or other contaminants. Drain and
replace if necessary. If bath water develops algae in less than
3 months make sure to clean the bath very thoroughly.
Every 6 Months
Replace bath water and clean the bath, regardless of
condition, to prevent algae growth.
Check the temperature and motor speed for accuracy.
Inspect the spindles for smooth rotation.
Annually
Inspect the drive belts and idler pulleys.
Inspect the O-rings, they should not be dried or cracked.
Every Two Years
Replace the drive belts.
Replace the O-rings.