Section J – Troubleshooting
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021 114 of 130
EAR99 technology subject to restrictions on cover page.
Mechanical Troubleshooting
Problem Possible Causes
Recommended
Action
Excessive
vibration in system
Belts are too tight or
too loose.
Replace or adjust
belts.
Tester feet are not
touching the bench.
Ensure all feet of the
tester are secure on
the bench.
Other vibration
sources.
Remove these other
sources or relocate
the tester.
Paddle/basket
heights not correct
Improper operation. Confirm that spindle
clamp and spindle are
properly engaged.
Confirm that paddle or
basket shaft is
properly tightened in
the spindle shaft.
Clamp has loosened or
spindle shaft was
replaced.
Perform the height
adjustment.