Section F – Operation
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021 74 of 130
EAR99 technology subject to restrictions on cover page.
Sampling Alarms tab
This screen provides the user with the ability to enable a
sampling alarm. The sampling alarm notifies the user when a
sample point is upcoming when a test is running. A message
is displayed on the screen with an alarm sound.
On/Off Switch: Enables/disables notifications at sampling
events.
Prenotification Time: The amount of time before a sample
point the alarm begins to sound. For example, if the user set
the alarm to 0:01, the alarm would begin 1 minute before each
sample point (0 – 9:59 in hh:mm).
NOTE: If enabled from the method, a sampling wizard appears
guiding the user when to take a manual sample.
Alarm Tone: The tone the instrument will play when the alarm
triggers. The current tone is played when the ‘Play’ icon is
touched.