Section G – Dissolution Testing
CD14 Tester & CD AutoPlus Autosampler User Guide—75-108-800 Rev. B—31 May 2021 84 of 130
EAR99 technology subject to restrictions on cover page.
Dissolution tab
The Dissolution tab provides information and settings for the
dissolution test.
Drive Speed: The target drive speed for the dissolution test.
The range is 0, 25 – 250 rpm.
Bath Temperature: The target bath temperature of the
dissolution tester. The range is 0, 25.00 - 51.00 °C.
Test Length: the total time of the dissolution test. The range is
0 – 99:00 (hh:mm).
Log Interval: The interval at which the tester status (speed,
temperature, etc.) is logged to the test report. The range is 0 to
99:00 (hh:mm). If the user does not wish the status to print, the
Print Interval should be set to 00:00.
NOTE: The tester status is printed by default at all sample
points, even if the print interval is set to 00:00.
Speed Tolerance: The tolerance for the motor speed for a
dissolution test. The default value is ± 4% of the set point. The
range is 0 (off), 1 – 10%.
Temperature Tolerance: The tolerance for the primary
temperature probe on the dissolution tester. The default value
is ± 0.5 °C. The range is 0 (off), 0.10 to 5.00 °C.
Digital Temperature Probes: When checked, allows vessel
temperature probes mounted on the AutoMag of the CD14
Tester to be monitored during the test.