4.4 Commands
PM100 / page 37
Reset Command
Command syntax:
*RST
Response syntax: None
Description: This command resets the device into a known state. It recalls
values from memory location #1 and enables the scrambler
unit. This is similar to the local operation mode (see also
IEEE488.2-1992-§10.32).
Service Request Enable Command
Command syntax:
*SRE <NR1>
Response syntax: None
Description: Sets the device’s Service Request Enable Register (see also
IEEE488.2-1992-§10.34 and Chapter ‘Status Reporting’).
Service Request Enable Query
Command syntax:
*SRE?
Response syntax:
<NR1 NUMERIC RESPONSE DATA>
Description: Queries the device’s Service Request Enable Register (see
also IEEE488.2-1992-§10.35 and Chapter ‘Status Reporting’).
Read Status Byte Query
Command syntax:
*STB?
Response syntax:
<NR1 NUMERIC RESPONSE DATA>
Description: Queries the device’s Status Byte (see also IEEE488.2-1992-
§10.36 and Chapter ‘Status Reporting’).
Standard Event Status Enable Command
Command syntax:
*ESE <NR1>
Response syntax: None
Description: Sets the device’s Standard Event Status Enable Register (see
also IEEE488.2-1992-§10.10 and Chapter ‘Status Reporting’).
Standard Event Status Enable Query
Command syntax:
*ESE?
Response syntax:
<NR1 NUMERIC RESPONSE DATA>