EasyManua.ls Logo

Tomey UD-800 - Page 120

Tomey UD-800
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-88
c) PC setting
Make connection settings and DICOM connection settings to export data to a personal
computer with TOMEY Link or Data Transfer installed. Touch the “DICOM Setting” button to
move to the DICOM setup screen (Fig. 2).
(1) Connection permission and selection of cable type
LAN
:
Use a LAN cable for connection.
When “LAN” is selected here, all items listed on this screen
need to be set.
USB
:
Use a USB cable for connection.
Set the Login User ID and Password. Items other than the
above do not need to be set.
Disable
:
Not connected with PC.
(2) Auto Inquiry
Set the auto inquiry reference for the patient's ID.
(3) Login Tomey Link
Set the user ID and password to log in to the TOMEY Link. Maximum
number of characters to be entered for the user ID or password is 16.
(4) Host IP address
Set the IP address of the PC to be connected. Touch the “DNS” button to
specify the PC name. A DNS server must be running in the LAN
environment to enable connection using the DNS.
Touch the input field to display the keyboard.
(5) Port number
A port number can be set.
Input range: 0 - 65535
Default setting: 80
(6) "Connection Test" button
Performs a connection test.
(7) DICOM - Settings
Moves to the DICOM settings screen.
(Fig. 1)
)
(Fig. 2)
)

Table of Contents