4-14
4.1.8 SRK/T Double K
This formula is for POST LASIK.
1. Implanted IOL power (D) for emmentropization
YCL
Xna
Pemme
11
1000
2. Implanted IOL power (D) for ametropia
post
post
amet
rCYVREFYCL
rLXVREFXna
P
1001.011
1001.01000
3. Predicted refractive power after surgery (D)
postpost
iol
rCYVCLPrLXVna
CLYPXna
REF
111001.01
111000
Where:
X = na•r
post
L1 (nc 1)
Y = na•r
post
C1 (nc 1)
L1 : Visual axial length (mm)
= L + (0.65696 0.02029L)
L : Axial length (mm)
C1 : Estimated anterior chamber depth after surgery (mm)
= H + Ofst.
Ofst: Calculated distance from the iris surface to the optical
surface of the implanted IOL (including corneal thickness)
(mm)
= ACD const 3.336
= (0.62467A 68.747) 3.336
H : Height of cornea dome (mm)
4/
22
Wrr
prepre
W : Calculated corneal diameter (mm)
= -5.41 + 0.58412LC + 0.098K
LC : Corrected axial length (mm)
L24.2 → LC = L
L > 24.2 → LC = 3.446 + 1.716L 0.0237L
2
A : A-Constant