3.1.4 Entering the patient data
When entering new patient information, carefully check that the same
ID is not registered in the database or other electronic medical records.
Only the first 14 digits of the ID number are displayed in the patient
information field. Check that the ID number is correct on the patient
information entry screen.
The network support system “TOMEY Link” (optional) is required to
use the patient information query function. “DATA Transfer” provided
with the instrument is not available for query of the patient data.
For details on the TOMEY Link Server settings, refer to the TOMEY Link
instruction manual.
Appropriate connection settings are required for connection with
TOMEY Link. Refer to "3.9.4 c) PC setting" to make settings.
Once examination data is output (printed, exported or saved), the ID
cannot be changed.
1) Touch the patient information button (1) to open the Patient Information
screen (Fig. 2).
- When both the patient information and examination data exist, a
warning message (Fig. 3) appears indicating you are about to
overwrite data when the patient information button is touched.
Touch the “Yes” button (7) to overwrite the data.