EasyManua.ls Logo

Tomey UD-800 - Holladay 1

Tomey UD-800
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-5
4.1.3 Holladay 1
1. Implanted IOL power (D)
SFCrYVREFYSFCL
rLXVREFXna
P
2001.022
2001.01000
2. Predicted refractive power after surgery (D)
SFCrYVQPrLXVna
YQPXna
REF
iol
2001.02
1000
Where:
X : nar L2 (nc 1)
Y : nar - (nc - 1)(C2 + SF)
Q : L2 C2 -SF
na : Aqueous humor and vitreous refractive index
= 1.336
nc : Corneal reflective index
= 4/3
L : Axial length (mm)
K : Average corneal refractive power (D)
= (K1 + K2)/2
r : Radius of corneal curvature (mm)
= {(KI - 1) × 1000}/K
KI : Cornea equivalent refractive index
SF : Distance from the iris surface to the optical center of the
implanted IOL (mm)
REF : Target Ref. after surgery (D)
V : Vertex distance (mm)
= 12
P : Implanted IOL power (D)
L2 : Corrected axial length (mm)
= L + 0.2
C2 : Anatomic anterior chamber depth; distance from the
corneal vertex to the iris surface (mm)
= 0.56 + Rag
4/
22
AGRag
r < 7 → Rag = 7
r 7 → Rag = r