3-92
When only this instrument and the Tomey Link Server or a
computer running Data Transfer are connected to the network,
the following settings can be used.
1) Check the computer IP address.
Check and record the IP address and subnet mask of the computer with DATA
Transfer installed. Refer to the DATA Transfer operation manual for details.
2) Settings on the Instrument
The settings shown in the table below describe an example when the IP address
of the computer with DATA Transfer installed is “192.168.2.128" and the subnet
mask is “255.255.255.0.” The IP setting method in this case is “Manual.”
Computer settings
Instrument settings
(Local)
IP address
192
168
2
128
192
168
2
129
Check these on the DATA
Transfer screen.
Same value as
computer setting
(*1)
Subnet
mask
255
255
255
0
255
255
255
0
Check these on the DATA
Transfer screen.
Same value as computer
setting
Default
gateway
-------
0
0
0
0
All “0”
Host IP
address
-------
192
168
2
128
IP address of computer
*1: An arbitrary number from 1 to 255 excluding numbers used by the computer